Findings from usage
Bottomed out on a surface, there are two magnifications which produce a crisp image around 50x and 230x
to get two these two magnifications, the side wheel has to be rolled.
LED at full brightness is required for anything non-luminous for quality reasons
angle of attack is non-important, so long as you can focus
steadiness of the microscope is important to the quality of image
this can be countered by not allowing the microscope to float
this can be done by pressuring the microscope on a sample
there is a capacitive button on the side which allows photos to be taken in the software
this is very annoying
Examples
Recommendations
add an actuator to control the focus
hard press the camera into samples to reduce unwanted movement